Required section · Section 5 of 6
Reconstruct a mistimed series
A second patient's workup is complicated by preanalytical and documentation problems. An ACTH tube was left at room temperature for several hours before it reached the laboratory. A cortisol was drawn with no recorded collection time. A late-night cortisol, drawn near midnight, arrived with no order note stating whether it was intended to test suppression or was simply a late random draw. None of the three values is missing from the analyzer's report; what is missing is the context each one needs to mean anything.
The laboratory professional's task is not to diagnose. It is to sort these three specimens into what is interpretable, what is compromised and must be flagged rather than reported as a normal or abnormal number, and what information has to travel with any result that is released. A mishandled or mistimed specimen is not the same thing as an abnormal result; treating it as an abnormal value, rather than as an unsuitable specimen, risks a clinician acting on a number the specimen never actually supported.
Releasing a paired ACTH/cortisol or a dynamic-test result without its timing and handling context can lead a clinician to misclassify adrenal reserve or HPA axis integrity, with downstream risk of over- or under-treating a patient who may be at risk for adrenal crisis. Protecting the specimen's metadata is part of protecting the patient, not a paperwork formality.
When a specimen's timing or handling breaks the analyte's requirement, flag it as unsuitable for interpretation and say why, rather than reporting a number the specimen cannot support.
Ordering exercise
Put this laboratory's ACTH plasma handling steps in the correct order under its validated controlled procedure; these requirements cannot be copied between systems.
1. Apply the local procedure's storage and transport step
Freeze or transport frozen only when the cited assay manufacturer, reference laboratory, or local controlled procedure requires it.
2. Use the local assay procedure
Apply only the cited assay manufacturer, reference-laboratory, or local controlled procedure.
3. Apply the local procedure's separation step
Use chilled centrifugation only when the cited assay manufacturer, reference laboratory, or local controlled procedure requires it.
4. Apply the local procedure's cooling step
Use ice only when the cited assay manufacturer, reference laboratory, or local controlled procedure requires it.
Knowledge checks
Reading and checks are open. Sign in only to save.
Knowledge check 1
Knowledge check 2
Knowledge check 3
Section status
Finish this section
Reading and checks are open. Sign in only to save.
The module finishes after every required section is marked done and every check in those sections is correct.