Module overview
Section 1 of 6 · Open sections

Required section · Section 1 of 6

A Verification Run That Almost Passed

You are reviewing a calibration verification run on the chemistry analyzer's glucose method. The method is a photometric glucose assay, reported in mg/dL, with an analytical measuring interval (AMI) of 0 to 400 mg/dL on this instrument. This laboratory's policy requires calibration verification (CV) at three levels, low, mid, and high, with each recovered value expected within plus or minus 10 percent of its target, consistent with the Clinical Laboratory Improvement Amendments (CLIA) requirement to check at least a minimal, a mid-point, and a high value near the top of the reportable range.

The low target is 40 mg/dL and recovers at 42 mg/dL, a 5 percent difference, within limit. The mid target is 200 mg/dL and recovers at 204 mg/dL, a 2 percent difference, within limit. The high target is 350 mg/dL and recovers at 390 mg/dL, an 11.4 percent difference, outside this laboratory's plus or minus 10 percent limit. Routine QC that morning, run at one normal-range level, had passed without comment.

The recovery chart below makes the pattern visible: the low and mid points sit inside the acceptance band around their targets, and the high point sits well above it. Nothing in the routine QC result predicted this, because routine QC does not test the top of the reportable range on this method.

Calibration, calibration verification, and QC are three separate, named activities with three separate purposes, and confusing them is how a drifted high end goes unnoticed.

A passing QC result tells you the levels QC actually tested are behaving, and nothing about the levels it did not test.

Illustrative drawing — this picture was drawn rather than captured.

Chart with three levels, low, mid, and high, each showing a target value, a light blue acceptance band of plus or minus 10 percent, and an observed recovery point. The low and mid points fall inside their bands. The high point, target 350 mg/dL, recovers at 390 mg/dL and is shown in coral, outside its band.
Figure 1Calibration verification recovery for the glucose method: low and mid levels recover inside the laboratory's acceptance band, the high level recovers outside it.
Glucose calibration verification: target, recovery, and acceptance against this laboratory's plus or minus 10 percent policy.
LevelTarget (mg/dL)Recovered (mg/dL)DifferenceWithin lab limit
Low4042+5.0%Yes
Mid200204+2.0%Yes
High350390+11.4%No

Knowledge checks

Reading and checks are open. Sign in only to save.

Knowledge check 1

Routine QC passed the morning of the glucose calibration verification failure. Why did QC not catch the high-level recovery problem?

Choose one option.

Section status

Finish this section

Reading and checks are open. Sign in only to save.

The module finishes after every required section is marked done and every check in those sections is correct.