Module overview
Section 4 of 6 · Open sections

Required section · Section 4 of 6

Working the Case: One Patient, Two Confirming Results

Return to the patient from the opening problem. Venous blood was drawn at 0730, fasting confirmed at 8 hours or more, into a properly filled gray-top sodium fluoride tube, mixed by inversion, and centrifuged and separated within 45 minutes of the draw per local turnaround target. Central-lab plasma glucose returned 131 mg/dL, above the fasting diabetes threshold of 126 mg/dL. Flag: high.

The same draw carried an A1c by an NGSP-certified HPLC method, with no variant flag on the chromatogram and no known hemoglobinopathy or anemia documented for this patient. The result was 7.1%, above the diagnostic threshold of 6.5%. Flag: high.

Two different tests, drawn the same day, both above their respective thresholds. That satisfies the two-abnormal-result rule without needing a repeat draw or a follow-up OGTT; the diagnosis is confirmed by same-day concordant results from two different tests, which is one of the two accepted confirmation pathways alongside the same test abnormal on two separate occasions.

Two more numbers were charted for this patient's visit and belong on the same problem list, but not in the diagnostic decision. A point-of-care capillary glucose from triage read 142 mg/dL, drawn nonfasting. It sits below the 200 mg/dL random threshold, this patient has no classic hyperglycemia symptoms to invoke the random-glucose pathway, and a point-of-care result is outside its validated diagnostic role regardless of the number. It is excluded from the decision, not deleted from the record.

When two different tests agree above threshold on the same day, the diagnosis is confirmed and no further testing is required for that purpose alone; document which specific results were used and why the others were not.

Illustrative drawing — this picture was drawn rather than captured.

Shared linear glucose axis showing 100, 126, 140, and 200 mg/dL at proportional heights. The patient’s fasting value is 131 mg/dL; the 2-hour sample was not performed.
Figure 1OGTT worksheet on one shared linear plasma-glucose axis: 100, 126, 140, and 200 mg/dL have proportional heights; the guided case fasting value is 131 mg/dL and no 2-hour specimen was performed.

Illustrative drawing — this picture was drawn rather than captured.

Flowchart showing fasting glucose of 131 and A1c of 7.1 percent both feeding into a box stating two different tests, same day, both above threshold, leading to diagnosis confirmed with no repeat testing required. A separate excluded branch shows a point-of-care glucose of 142 and an A1c of 5.9 percent with a sickle trait variant flag, both marked as excluded from the decision.
Figure 2Workflow from two same-day abnormal results to a confirmed diagnosis, with excluded results held apart.
Results charted for the guided case, same visit.
TestResultThreshold usedUsed for diagnosis?
Fasting plasma glucose (central lab, 0730 draw)131 mg/dL≥126 mg/dL diabetesYes, one of two confirming results
A1c (NGSP-certified HPLC, same draw)7.1%≥6.5% diabetesYes, one of two confirming results
Point-of-care glucose (triage, nonfasting)142 mg/dL≥200 mg/dL random+symptomsNo, outside validated diagnostic role

Knowledge checks

Reading and checks are open. Sign in only to save.

Knowledge check 1

The guided patient has a same-day fasting glucose of 131 mg/dL and A1c of 7.1%, both above threshold. What is the correct next step for confirming the diagnosis?

Choose one option.

Section status

Finish this section

Reading and checks are open. Sign in only to save.

The module finishes after every required section is marked done and every check in those sections is correct.