Module overview
Section 4 of 6 · Open sections

Required section · Section 4 of 6

Working the calibration line: a glucose result

This example uses a reference-type enzymatic glucose method, hexokinase / glucose-6-phosphate dehydrogenase (hexokinase-G6PD); it is not tied to one named commercial instrument. Glucose is phosphorylated by hexokinase using ATP, and the resulting glucose-6-phosphate is oxidized while NADP+ is reduced to NADPH. NADPH absorbs light at 340 nm, and the rise in absorbance, delta A, is proportional to glucose concentration within the method's linear range. This is an endpoint method: the analyzer waits for the reaction to reach a stable plateau and reads delta A once, rather than tracking a rate.

Three calibrators establish the signal-to-concentration relationship for this run: a zero calibrator, a mid-level calibrator, and a high-level calibrator, each with an assigned glucose concentration. Their delta A values at 340 nm are 0.000 at 0 mg/dL, 0.300 at 100 mg/dL, and 0.900 at 300 mg/dL. Plotted against concentration, these three points fall on a straight line with a slope of 0.003 delta A per mg/dL, which is the calibration curve the analyzer stores and applies to every patient specimen run in this batch.

The patient specimen in this case produces a final delta A of 0.546 once the reaction plateaus, shown as the coral point on the endpoint curve below. Applying the calibration slope, 0.546 divided by 0.003 delta A per mg/dL, gives a reported glucose of 182 mg/dL. Two internal-consistency checks matter here. First, units are mg/dL throughout, calibrator and patient alike. Second, the patient's delta A of 0.546 falls between the mid and high calibrator readings, so the analyzer is interpolating within its calibrated range rather than extrapolating past the top calibrator, which is a more defensible position for the result to sit in.

The right-hand panel of the same figure shows a kinetic reaction on the same time axis for contrast, using a different analyte. Instead of waiting for a plateau, the analyzer reads absorbance repeatedly and calculates the rate of change, delta A per minute, over the segment where the reaction proceeds linearly. Neither approach is more correct in general; each is chosen for the chemistry of a specific method, endpoint for many concentration assays and kinetic for most enzyme activities.

This glucose value of 182 mg/dL sits inside an AMR of 10 to 500 mg/dL, so no dilution or rerun is needed before it can be reported. A fasting reference interval of roughly 70 to 99 mg/dL is shown for illustration only; the actual interval a laboratory reports is method- and population-specific and comes from that laboratory's own verified reference interval.

A reported concentration is only as trustworthy as the calibration line it was read against, so before you accept a number, know whether the patient's signal fell inside the calibrated range or had to be extrapolated beyond it.

Illustrative drawing — this picture was drawn rather than captured.

Two line charts on the same time axis: the left chart rises quickly then plateaus at delta A 0.546 around 150 seconds, with the endpoint read marked in coral; the right chart rises in a steady straight line, with the rate-calculation window between two coral points marked by a teal dashed reference line.
Figure 1Endpoint reaction reaching a stable plateau at delta A 0.546 compared with a kinetic reaction rising at a steady rate
Absorbance readings for an endpoint hexokinase glucose reaction and a kinetic reaction on the same time axis
Time (seconds)Endpoint reaction, delta A at 340 nmKinetic reaction, delta A at 340 nm
00.0000.050
300.1500.110
600.4000.170
900.5200.230
1200.5450.290
1500.546 (read taken)0.350

Knowledge checks

Reading and checks are open. Sign in only to save.

Knowledge check 1

The patient's delta A of 0.546 falls between the mid calibrator's 0.300 (100 mg/dL) and the high calibrator's 0.900 (300 mg/dL). What does this tell you about the reported glucose of 182 mg/dL?

Choose one option.

Knowledge check 2

Which statement correctly separates a measured analyte from a calculated one, using creatinine and eGFR as the example?

Choose one option.

Section status

Finish this section

Reading and checks are open. Sign in only to save.

The module finishes after every required section is marked done and every check in those sections is correct.