Module overview
Section 5 of 6 · Open sections

Required section · Section 5 of 6

The potassium result: where do you look first?

Return to the opening case. The patient's potassium result is 6.8 mmol/L. QC is in range for both levels, calibration on the potassium electrode is current, and the instrument raised no error flag on this specimen. The event log below lays out the run in the order it happened, so the 'no QC failure' premise is something you can read directly rather than take on faith.

The collection notes matter as much as the analyzer output. This specimen was drawn with a small-gauge needle after a difficult venipuncture, and there was a delay before it reached the laboratory. Both are preanalytical circumstances, meaning they happened before the specimen ever reached the analyzer, and both are recognized mechanisms for hemolysis: rupture of red cells that releases their high intracellular potassium content into the serum or plasma, producing a falsely elevated potassium result that has nothing to do with the patient's true blood chemistry. This is distinct from in-vivo hemolysis, a real patient process that can itself cause true hyperkalemia; distinguishing the two is a laboratory and clinical judgment made from history, appearance of the specimen, and other findings, not a rule that settles the question in the abstract.

Two paths do not answer the question. Rerunning quality control will not help; QC already passed, and QC tests the system, not this specimen's integrity. A formula-based correction of the potassium value using a hemolysis index is not considered reliably accurate and should not generally substitute for recollection and repeat testing when the patient is clinically stable enough to allow it; mathematically adjusting the number is not the same as knowing the true value.

What does answer the question is checking specimen integrity first: was this specimen visibly hemolyzed, and does the collection history support a preanalytical explanation. If hemolysis is the leading, testable explanation, the defensible action is to request or collect a fresh specimen, collected without the same difficulty, and repeat the potassium measurement before treating 6.8 mmol/L as a true reflection of this patient's blood chemistry. This action is preferred because it produces a directly comparable, trustworthy number rather than an estimate.

Illustrative drawing — this picture was drawn rather than captured.

A six-row event log listing specimen collection with a difficult draw, receipt with a transport delay, QC pass for both levels, current calibration, a potassium result of 6.8 mmol/L with no error flag highlighted in coral, and an autoverification hold for manual review.
Figure 1Run record showing collection notes, QC pass, current calibration, a potassium result with no error flag, and an autoverification hold for manual review.
Timeline of events for the potassium case
TimeEventStatus
07:42Specimen collected, difficult venipuncture, small-gauge needlelogged
08:15Specimen received in laboratory, transport delay notedlogged
08:20QC level 1 and level 2, chemistry panelpass
08:20Calibration status, potassium electrodecurrent
08:24Potassium result 6.8 mmol/Lno error flag
08:24Autoverificationheld for manual review

Knowledge checks

Reading and checks are open. Sign in only to save.

Knowledge check 1

QC passed, calibration is current, and no instrument error flag fired, yet the potassium result is 6.8 mmol/L with a difficult draw and a transport delay in the history. Where should you look first?

Choose one option.

Knowledge check 2

Given the same case, select every action that is defensible laboratory practice here.

Choose at least 2 options.

Section status

Finish this section

Reading and checks are open. Sign in only to save.

The module finishes after every required section is marked done and every check in those sections is correct.