Module overview
Section 1 of 6 · Open sections

Required section · Section 1 of 6

A Level 2 control flags high after a reagent lot change

At 07:00 the satellite laboratory runs its assigned external controls on a nonwaived quantitative whole-blood assay. Level 1 targets 2.00 units, acceptable 1.90 to 2.10 units. Level 2 targets 8.00 units, acceptable 7.60 to 8.40 units. Both controls are pulled from refrigerator storage, treated exactly like a patient specimen, and accepted at 07:04. The analyzer's electronic internal check passes before each cycle, same as every morning.

At 10:18 a new reagent lot goes on the analyzer. The courier record for that shipment shows a documented temperature excursion in transit. At 10:25 the operator reruns controls before resuming patient testing. Level 1 reads 2.04 units, inside its range. Level 2 reads 8.63 units, above its 8.40 unit upper limit and flagged high.

The internal electronic check passed both times. Nothing about the analyzer's self-check caught this. Only the external Level 2 control, tested the same way a patient specimen is tested, caught the shift. That gap between what an internal check can see and what a full-process control can see is the reason a laboratory builds a quality control plan around where a failure can actually occur, not around running one type of check by habit.

A structured risk assessment is how a laboratory designs quality control that has a real chance of catching a failure like this one, instead of assuming the manufacturer's minimum or the last laboratory's plan will do it. A control that never fails to detect anything is not proof the process is safe; it may only mean the control was never placed where the actual risk was.

Knowledge checks

This section has no knowledge checks.

Section status

Finish this section

Reading and checks are open. Sign in only to save.

The module finishes after every required section is marked done and every check in those sections is correct.