Photometry and atomic absorption
15 min
- Trace a photometric fault to the lamp, wavelength selector, cuvette, or detector
- Predict the effect of stray light on a high-absorbance measurement
- Calculate a concentration from absorbance with the fitted standard curve
- Match an atomic-absorption interference to its mechanism
Try first
Get the idea
Follow the light
A photometer sends light from a source through a wavelength selector and a sample cell to a detector. A signal processor turns the detector's response into a result.1,2 Each symptom points to the part whose job it affects.
| Part | Its job | Failure clue |
|---|---|---|
| Source | Supplies light across the working band | Low or unstable energy |
| Wavelength selector | Isolates the intended band | Wrong peak position, excess bandwidth, stray light |
| Sample cell | Holds the specimen at a set path length | Scratches, bubbles, unequal path length |
| Detector | Converts light into an electrical signal | Noise, saturation, loss of response |
| Signal processor | Applies blanking and calibration | Wrong factor, unit or curve fit |
Ordinary glass and many plastics absorb ultraviolet light. A 340 nm reading of NADH needs a cell of quartz, fused silica or another material that transmits ultraviolet light.1
Stray light pulls high absorbance down
Stray light is off-band light that reaches the detector from scatter, dust, scratched optics or higher grating orders. At high absorbance little light passes the sample, so the added light is a large share of what the detector receives. Transmittance reads too high and absorbance too low. With 1% stray light, a true absorbance of 2.00 reads about 1.70.1,2 A cutoff filter or a strongly absorbing solution that should pass no light at the test wavelength detects stray light. Holmium oxide or didymium glass checks wavelength accuracy.1
Read the concentration off the fitted line
A validated straight line with an intercept gives the concentration as:
C = (A − intercept) ÷ slope
The single-point ratio, (specimen absorbance ÷ calibrator absorbance) × calibrator concentration, assumes a straight line through the origin. On a line with an intercept, it biases results most at concentrations far from the calibrator.1,3 Read results only inside the validated interval.3
Match the atomic absorption correction to the cause
Atomic absorption measures how much of the lamp's element line free ground-state atoms remove in a flame or graphite furnace.1
- Chemical. Phosphate binds calcium in a compound that resists atomization. A releasing agent such as lanthanum frees the calcium.
- Ionization. A hot flame ionizes the atoms. An ionization buffer or a cooler flame corrects it.
- Background. The matrix absorbs and scatters light at the line. Background correction removes that signal.
- Physical. Viscosity or dissolved solids change nebulization. Matrix matching, dilution or standard addition corrects it.1,2
References
- Bishop ML, Fody EP, Van Siclen C, Mistler JM, Moy M. Clinical Chemistry: Principles, Techniques, and Correlations. 9th ed. Jones & Bartlett Learning; 2023.
- Rifai N, Chiu RWK, Young I, Burnham CAD, Wittwer CT, eds. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023.
- Clinical and Laboratory Standards Institute. Evaluation of Linearity of Quantitative Measurement Procedures. 2nd ed. CLSI guideline EP06. Clinical and Laboratory Standards Institute; 2020. Accessed September 26, 2026. https://clsi.org/shop/standards/ep06/
Watch one
A colorimetric method's validated calibration is a straight line fitted to calibrators from 25 to 400 mg/dL:
A = (0.00250 dL/mg) × C + 0.050
The 200 mg/dL calibrator reads 0.550, and a specimen reads 0.300. What concentration does the line give for the specimen?
- Note the model: slope 0.00250 dL/mg, intercept 0.050. The calibrator fits it, since 0.00250 × 200 + 0.050 = 0.550.
The validated model sets the calculation, and this line has an intercept.
- Subtract the intercept: 0.300 − 0.050 = 0.250.
The intercept is signal present at zero concentration, so it comes off before dividing.
- Divide by the slope: 0.250 ÷ 0.00250 dL/mg = 100 mg/dL.
The slope is absorbance per unit of concentration, so dividing by it returns the concentration.
- Check the shortcut: (0.300 ÷ 0.550) × 200 mg/dL = 109 mg/dL, 9% higher than the line's result.
Comparing with the ratio shows the size of the bias the shortcut would add.
- Check the interval: 100 mg/dL lies between 25 and 400 mg/dL.
A result is read directly only inside the validated interval.
Your turn
Use it
- The chemistry analyzer's scheduled photometer checks run at 340 nm before the day shift.
- Lamp energy is within the manufacturer's limits.
- The holmium oxide wavelength check passes.
- Linearity standards read, as expected absorbance to measured absorbance: 0.500 to 0.495, 1.000 to 0.981, 1.500 to 1.438, 2.000 to 1.826.
- Both control levels are within limits.
- Specimens waiting for testing include several with high values.
The clue that settles this case is the shape of the linearity error:
- The 0.500 standard reads within 1% of expected.
- The 2.000 standard reads about 9% low.
- Lamp energy and wavelength position both pass.
An error that grows with absorbance, with the lamp and the peak position in order, points to stray light in the wavelength selector's optics. The cutoff solution confirms it, and high specimens are tested on a photometer that passes its checks.
Results
- Trace a photometric fault to the lamp, wavelength selector, cuvette, or detector
- Predict the effect of stray light on a high-absorbance measurement
- Calculate a concentration from absorbance with the fitted standard curve
- Match an atomic-absorption interference to its mechanism
To review
6 questions from this step will come back in Review.
Next step: Chromatography, electrophoresis, mass spectrometry and osmometryReview nowOpen the part
Keep
Sources checked
The rest of this step
A short briefing, a demonstration at the bench, 3 practice problems and a short case.
A free account opens the rest and keeps your progress.