Pipettes, glassware and temperature checks
15 min
- Verify temperature at the working sample position
- Follow the to-contain or to-deliver marking on volumetric equipment
- Choose a pipette type for the volume and for viscous, volatile, or foaming liquids
- Calculate delivered volume from water mass and the stated conversion factor
Try first
Get the idea
Measure where the specimen sits
A water bath or heating block display reports the controller's own sensor. The temperature at the specimen can differ with position, gradients and drift, so the check is made at the sample position.1
- Let the bath and a calibrated reference probe equilibrate.
- Read the probe at the working setpoint in representative rack positions.
- Apply the probe's calibration correction.
- Compare each corrected reading with the procedure's acceptance interval.1
The probe's calibration record names the reference device, the correction factors and their uncertainty. A "NIST traceable" label by itself is a claim.2
Read the marking before you drain
To-contain (TC) ware holds its stated volume. To-deliver (TD) ware releases its stated volume when its drainage procedure is followed.3
- A continuous etched band or double ring near the top marks a blowout pipette. Its last drop is expelled.
- A pipette without that mark is self-draining. The drop left in its tip is part of its calibration.1
Using one drainage rule for every pipette adds or leaves behind the same volume on every transfer.
Match the pipette to the liquid
An air-displacement pipette moves liquid through an air cushion. Viscosity, volatility, foaming and temperature differences change what that cushion delivers. A positive-displacement pipette moves its piston in contact with the liquid and handles these liquids better. Reverse pipetting is another option, and either change is verified locally. The pipette's operating range must cover the volume.1
Weigh water to find the volume
Gravimetric calibration weighs replicate deliveries of purified water. The factor Z converts mass to volume and folds in water density and air buoyancy.4
mean delivered volume = mean mass × Z
systematic error = mean delivered volume − selected volume
At 22.0 °C and 101.3 kPa, Z is 1.0033 µL/mg.4 A 100 µL pipette with a mean mass of 99.29 mg delivers 99.29 mg × 1.0033 µL/mg = 99.62 µL, a systematic error of −0.38 µL. Reading the milligrams as microliters overstates the error as −0.71 µL. The laboratory compares the systematic and random errors with its acceptance limits.5
References
- Bishop ML, Fody EP, Van Siclen C, Mistler JM, Moy M. Clinical Chemistry: Principles, Techniques, and Correlations. 9th ed. Jones & Bartlett Learning; 2023.
- Bruce SS, Possolo A, Watters RL Jr. Metrological Traceability: Frequently Asked Questions and NIST Policy. NIST Technical Note 2156. National Institute of Standards and Technology; 2021. Accessed September 26, 2026.
- ASTM International. Standard Specification for Laboratory Glass Volumetric Apparatus. ASTM E694-18(2024). ASTM International; 2024.
- International Organization for Standardization. Piston-operated volumetric apparatus — Part 6: Gravimetric reference measurement procedure for the determination of volume. ISO 8655-6:2022. International Organization for Standardization; 2022. Accessed September 26, 2026.
- International Organization for Standardization. Piston-operated volumetric apparatus — Part 2: Pipettes. ISO 8655-2:2022. International Organization for Standardization; 2022. Accessed September 26, 2026.
Watch one
You check a 500 µL air-displacement pipette at its nominal setting.
- Ten deliveries of purified water weigh 4,974.0 mg in total.
- The water and air are at 22.0 °C and the pressure is 101.3 kPa, so Z is 1.0033 µL/mg.
- The laboratory's procedure accepts a systematic error within ±4.0 µL at this setting.
Does the pipette pass for systematic error?
- Find the mean mass: 4,974.0 mg ÷ 10 = 497.40 mg.
The check judges the pipette on the average of its replicate deliveries.
- Convert to volume: 497.40 mg × 1.0033 µL/mg = 499.04 µL.
Z corrects for water density and air buoyancy, so each milligram on the balance stands for slightly more than 1 µL.
- Subtract the selected volume: 499.04 µL − 500 µL = −0.96 µL, which is −0.19% of the setting.
The sign shows the direction, and a negative error means the pipette delivers too little.
- Compare with the limit: −0.96 µL lies within ±4.0 µL.
The procedure's limit decides pass or fail for this setting.
Your turn
Use it
- You set up a manual enzyme immunoassay on the special chemistry bench.
- Step 1 adds 20 µL of a viscous, glycerol-based conjugate to each tube.
- Step 2 incubates the tubes for 30 min in a heating block set to 37 °C. The procedure accepts 36.0 to 38.0 °C in every well used.
- Step 3 adds 1.0 mL of stop solution with a 1 mL serologic pipette marked TD, with a double ring near the mouthpiece.
- This morning's check with a calibrated reference probe, correction applied, read 37.1 °C in a center well and 35.6 °C in a corner well. The display reads 37.0 °C.
Each item in this setup showed what it needed through something you could check:
- The conjugate's viscosity chose the positive-displacement pipette.
- The corner well's corrected reading of 35.6 °C, below the 36.0 °C limit, settled the heating block. The display and the center well could not show it.
- The double ring near the mouthpiece made the serologic pipette a blowout pipette.
Results
- Verify temperature at the working sample position
- Follow the to-contain or to-deliver marking on volumetric equipment
- Choose a pipette type for the volume and for viscous, volatile, or foaming liquids
- Calculate delivered volume from water mass and the stated conversion factor
To review
6 questions from this step will come back in Review.
Next step: Water, vessels, balances and centrifugesReview nowOpen the part
Keep
Sources checked
The rest of this step
A short briefing, a demonstration at the bench, 3 practice problems and a short case.
A free account opens the rest and keeps your progress.