After QC fails
14 min
- Confirm corrective action before accepting the next QC run
- Decide which patient results need review after rejected QC
- Set a new control lot's mean from its crossover runs and its SD from the established CV
Try first
Get the idea
A rejected QC run means the analyzer cannot be trusted for that analyte until someone shows otherwise. Reporting of affected patient results stops, and three pieces of work follow:
- Correct the fault.
- Prove the correction.
- Look back at what was already reported.
Correct and prove
The investigation points at a cause, such as:
- a control vial made up wrongly
- a reagent pack past its onboard stability
- a drifted calibration
- a clogged probe
The technologist fixes that cause and then runs QC again. Acceptable QC after a named correction is what returns the test to service.
Rerunning the same controls until they pass skips the first half. The rerun may land inside the limits by chance while the fault is still present, and the next run fails the same way.
CLIA requires the laboratory to document every corrective action it takes when control or calibration results fail its acceptance criteria, so the record says what went wrong and what was done.1
Look back
- Patient results obtained in the unacceptable run, and all results reported since the last acceptable run, are evaluated to see whether they were affected.1
- The window reaches back that far because a shift or trend can start before the run that finally breaks a rule.
- Following the laboratory's procedure, specimens from the window are retested after the correction.
- When a reported result proves to be in error, the laboratory notifies the ordering provider and issues a corrected report promptly.3
A new control lot
Recovery sometimes means opening a new lot, and every lot change needs its own limits.
Take the target mean from the new lot itself:
- The new lot has its own concentration, so its target mean comes from its own crossover runs, tested in parallel with the current lot. About 10 measurements on separate days give a starting target.2
- The old lot's mean would make every new-lot result look shifted.
- The insert mean is assigned across many laboratories and can sit away from this analyzer's own center.
Set the SD:
- A few days of crossover runs show only short-term scatter and give limits too tight to hold.
- When the CV is steady across concentrations, the established long-term CV carries over:2
new SD = new mean × established CV ÷ 100
The mean and SD are then updated as stable data accumulate.
References
- Electronic Code of Federal Regulations. 42 CFR §493.1282: Standard: Corrective actions. Accessed September 23, 2026.
- Clinical and Laboratory Standards Institute. Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. 4th ed. CLSI guideline C24Ed4E. Clinical and Laboratory Standards Institute; 2016. Accessed September 23, 2026.
- Electronic Code of Federal Regulations. 42 CFR §493.1291: Standard: Test report. Accessed September 23, 2026.
Watch one
At 07:00 both alanine aminotransferase (ALT) control levels read below −2 SD, breaking the 2-2s rule.
- The last acceptable QC was at 15:00 yesterday.
- The reagent pack on board was opened 31 days ago, and its onboard stability is 28 days.
What has to happen before ALT results are reported again?
- Stop reporting ALT and keep the run's patient results pending release.
Results from a rejected run cannot go out while their accuracy is in question.
- Take the expired reagent pack off and load a fresh pack.
Both levels moving the same way points to a systematic cause, and a pack past its onboard stability is one.
- Run both control levels on the fresh pack and confirm they are inside their limits.
Only acceptable control results show that the new pack fixed the fault.
- Document the rejection, the expired pack, the replacement and the acceptable QC.
The record has to name the cause and the fix, so the recovery can be checked later.
- Evaluate the patient ALT results from the 07:00 run and every ALT reported since 15:00 yesterday.
A drift in reagent can start before the run that finally fails.
Your turn
Use it
Results
- Confirm corrective action before accepting the next QC run
- Decide which patient results need review after rejected QC
- Set a new control lot's mean from its crossover runs and its SD from the established CV
To review
3 questions from this step will come back in Review.
Next step: How well a method performsReview nowOpen the part
Keep
Sources checked
The rest of this step
A short briefing, a demonstration at the bench, 3 practice problems and a case in the lab.
A free account opens the rest and keeps your progress.