The testing cycle and patient identity
12 min
- Locate an error in the preanalytic, analytic, or postanalytic phase
- Verify patient and specimen identity before testing or giving a result
Try first
Get the idea
Three phases
Every result passes through three connected phases, often called preanalytic, analytic and postanalytic. ISO standards name them preexamination, examination and postexamination. Together they make up the path of workflow, which starts with the patient and ends when the result is reported and interpreted.1
| Phase | Covers | Typical failures |
|---|---|---|
| Preanalytic | Test selection, patient identification, collection, labeling, transport and receipt | Wrong patient, wrong tube, short fill, clot, hemolysis, delay |
| Analytic | The measurement itself | Failed quality control, calibration error, reagent or instrument fault |
| Postanalytic | Review, reporting and interpretation | Transcription error, wrong reference interval, late critical call |
Name the phase where it began
Naming the phase tells you where to look. A problem often shows up in a later phase than the one where it began.
- Hemolysis is the usual example. The analyzer's index reports it during measurement, and in most cases the red cells broke during collection or transport, so changing reagents or recalibrating fixes nothing.1,2
- Acceptable quality control shows the measuring system worked. It says nothing about whether one patient's tube was fit to measure.1
Identity at every handoff
The first preanalytic check is identity, and it runs through the whole cycle.
- A specimen is linked to a patient by at least two person-specific identifiers, such as full name with date of birth or medical record number.3,4
- A room or bed number never counts, because patients move and rooms are reused.4
- The label goes on the tube in the patient's presence, with the name, a unique identification number, the collection date and time and the collector's initials.1,3
The same two identifiers are checked again each time the specimen or its result changes hands:3,4
- at receipt, against the order
- before testing
- whenever a result is given out, including by phone
A name alone is one identifier, and two patients in the same hospital can share it. An unlabeled specimen, or one whose label and request do not match, is rejected under the laboratory's identification policy. The requester is told promptly, and a new specimen is requested.1
References
- World Health Organization. Laboratory Quality Management System: Handbook. World Health Organization; 2011:11,65-66. Accessed September 23, 2026.
- Simundic AM, Bölenius K, Cadamuro J, et al. Joint EFLM-COLABIOCLI recommendation for venous blood sampling. Clin Chem Lab Med. 2018;56(12):2015-2038. doi:10.1515/cclm-2018-0602
- Clinical and Laboratory Standards Institute. Patient and Laboratory Specimen Identification Processes. 1st ed. CLSI standard PRE01. Clinical and Laboratory Standards Institute; 2024. Accessed September 23, 2026. https://clsi.org/shop/standards/pre01/
- The Joint Commission. Two patient identifiers: understanding the requirements. Updated April 21, 2026. Accessed August 28, 2026.
Watch one
An outpatient serum collected at a clinic at 09:00 reaches the laboratory unspun at 15:30.
- Potassium is high and glucose is low.
- The hemolysis index is not raised.
- Both control levels on the run were acceptable, and calibration is current.
Where did the error begin?
- Check the analytic evidence: both controls acceptable and calibration current, so the measuring system worked on this run.
The run's QC and calibration records settle the analytic question quickly.
- Check the indices: hemolysis is not raised, so hemolysis does not explain the potassium.
A raised potassium on red serum would point to broken red cells.
- Read the collection and receipt times: six and a half hours with serum sitting on its cells.
Time and handling records show what happened to the tube before it was measured.
- Match the pattern: a high potassium with a low glucose fits delayed separation.
Cells left in contact with serum keep using glucose and let potassium leak out.
- The error arose before measurement: preanalytic. Apply the laboratory's stability limit and request a new specimen.
The phase is named for where the error began, which is also where the fix belongs.
Your turn
Use it
Results
- Locate an error in the preanalytic, analytic, or postanalytic phase
- Verify patient and specimen identity before testing or giving a result
To review
3 questions from this step will come back in Review.
Next step: Is the specimen fit to test?Review nowOpen the part
Keep
Sources checked
The rest of this step
A short briefing, a demonstration at the bench, 3 practice problems and a case in the lab.
A free account opens the rest and keeps your progress.