After QC fails

14 min

Skills you'll build3 skills
  • Confirm corrective action before accepting the next QC run
  • Decide which patient results need review after rejected QC
  • Set a new control lot's mean from its crossover runs and its SD from the established CV

At the bench: Review QC before releasing results

Try first

Try first

A glucose run is rejected when both control levels fall below −2 SD. Without changing anything, you rerun both controls, and this time both are inside their limits. What do you do?

Keep going to see the reasoning.

Keep going to see the reasoning.

Right. Keep going to follow the reasoning.

Keep going to see the reasoning.

Get the idea

After a rejected QC run, patient results for that analyte are not reported until the problem is corrected. Three steps follow:

  1. Correct the fault.
  2. Confirm the correction with QC.
  3. Review the results already reported.

Correct the cause and confirm with QC

The investigation identifies a cause, such as:

  • a control vial reconstituted incorrectly
  • a reagent pack past its onboard stability
  • a drifted calibration
  • a clogged probe

The laboratory professional corrects that cause and then runs QC again. The test returns to service only after acceptable QC following a documented correction.

Rerunning the same controls until they pass skips the correction. The rerun may fall within the limits by chance with the fault still present, and the next run fails the same way.

CLIA requires the laboratory to document every corrective action it takes when control or calibration results fail its acceptance criteria, so the record shows what went wrong and what was done.1

Patient result lookback

  • Patient results obtained in the unacceptable run, and all results reported since the last acceptable run, are evaluated to see whether they were affected.1
  • The review goes back that far because a shift or trend can start before the run that violates a rule.
  • Following the laboratory's procedure, specimens from the window are retested after the correction.
  • When a reported result proves to be in error, the laboratory notifies the ordering provider and issues a corrected report promptly.3

A new control lot

Correction sometimes means opening a new control lot, and every new lot needs its own limits.

Establish the target mean from the new lot:

  • The new lot has its own concentration, so its target mean comes from its own crossover runs, tested in parallel with the current lot. About 10 measurements on separate days give a starting target.2
  • The old lot's mean would make every new-lot result look shifted.
  • The insert mean is assigned across many laboratories and can differ from this analyzer's own mean.

Set the SD:

  • A few days of crossover runs show only short-term scatter and give limits that are too tight.
  • When the CV is steady across concentrations, the established long-term CV is applied to the new lot:2

new SD = new mean × established CV ÷ 100

The mean and SD are then updated as stable data accumulate.

Follow the bench demonstration

Sources for this step
References
  1. Electronic Code of Federal Regulations. 42 CFR §493.1282: Standard: Corrective actions. Accessed September 23, 2026.
  2. Clinical and Laboratory Standards Institute. Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. 4th ed. CLSI guideline C24Ed4E. Clinical and Laboratory Standards Institute; 2016. Accessed September 23, 2026.
  3. Electronic Code of Federal Regulations. 42 CFR §493.1291: Standard: Test report. Accessed September 23, 2026.

Watch one

At 07:00 both alanine aminotransferase (ALT) control levels read below −2 SD, violating the 2-2s rule.

  • The last acceptable QC was at 15:00 yesterday.
  • The reagent pack on board was opened 31 days ago, and its onboard stability is 28 days.

What has to happen before ALT results are reported again?

  1. Stop reporting ALT and keep the run's patient results pending release.

    Results from a rejected run cannot be reported while their accuracy is in question.

  2. Take the expired reagent pack off and load a fresh pack.

    Both levels moving the same way suggests a systematic cause, and a pack past its onboard stability is one.

  3. Run both control levels on the fresh pack and confirm they are inside their limits.

    Only acceptable control results show that the new pack corrected the problem.

  4. Document the rejection, the expired pack, the replacement and the acceptable QC.

    The record names the cause and the correction so it can be reviewed later.

  5. Evaluate the patient ALT results from the 07:00 run and every ALT reported since 15:00 yesterday.

    A drift in reagent can start before the run that finally fails.

Replace the expired pack, obtain acceptable QC, document both, then resume reporting and review every ALT result back to the last acceptable run at 15:00 yesterday.

Practice this skill

Your turn

Problem 1 of 3 · Hints available

Potassium QC was acceptable at 10:00. Patient potassium results were reported through the late morning. At 14:00 a control violates the 1-3s rule. After the fault is corrected, which patient results do you evaluate?

That covers the unacceptable run and everything since the last acceptable run, which is the whole stretch where the fault could have been present.

QC at 10:00 was acceptable, so it supports the results reported before it. The review starts at the last acceptable run.

The fault may have been present before the run that caught it, so results reported after 10:00 can be affected too.

Reviewed only the failed run's patient results

A shift or trend can begin before the run that fails. The review covers patient results from the unacceptable run and every result reported since the last acceptable run, so reviewing the failed run alone can leave affected results uncorrected in the record.

Hint
  1. The rule was violated at 14:00, and the fault may have started earlier.
  2. Find the last acceptable QC run.

Review Resolving rejected QC

Problem 2 of 3 · Hints available

A new cholesterol control lot ran 10 times over 10 days beside the current lot. Its crossover mean is 212 mg/dL, and the SD of those 10 runs is 1.9 mg/dL. The insert lists a mean of 205 mg/dL with an SD of 8 mg/dL. The established CV for this level is 2.0%. Which starting limits do you set?

The insert values are assigned across many laboratories. Its mean differs from this analyzer's own mean, and its wide SD lets a real shift continue longer before it is detected.

Set the provisional mean from the old lot or the insert

A new control lot has its own concentration, so every new-lot result plotted against the old lot's mean looks like a shift. The package insert mean is assigned from many laboratories and can differ from the mean on this analyzer. Establish the mean from the new lot's own crossover runs.

The mean comes from the lot's own crossover runs, and the SD is 212 × 2.0 ÷ 100 = 4.24 mg/dL, about 4.2 mg/dL.

The mean fits this lot. An SD of 1.9 mg/dL reflects only 10 days of short-term scatter, so limits built on it are too tight and will reject good runs.

Set the provisional SD from the crossover runs or the insert

A short crossover study captures only short-term imprecision and gives limits that are too tight. The package insert SD covers many laboratories, so a real shift goes undetected longer. Apply the established CV to the new mean to get an SD that includes day-to-day, calibration, and reagent lot variation.

Hint
  1. The mean has to describe this lot on this analyzer.
  2. The SD should include day-to-day, calibration and reagent variation, which a 10-day study does not fully capture.
  3. Apply the established CV to the new mean.

Review Control materials

Problem 3 of 3 · On your own

A creatinine run is rejected under the R-4s rule. The laboratory professional finds a small clot on the sample probe and clears it. What returns creatinine to patient testing?

The clot was a likely cause, and acceptable QC afterward shows whether the analyzer is working. Clearing the clot is not confirmed as the correction until controls are acceptable.

Repeating controls until most pass leaves the outcome to chance. The correction is confirmed by one acceptable QC run after the fix, documented with it.

Accepted repeat QC without a documented correction

A rerun control can fall in range by chance, and repeating controls until they pass does not find the cause. Patient testing then resumes on a system that may fail again. Identify and correct the cause, document the corrective action, and confirm it with acceptable QC before resuming patient testing.

A named correction followed by acceptable QC, both on record, returns the test to service. The lookback to the last acceptable run follows.

Review Resolving rejected QC