After QC fails
14 min
Skills you'll build3 skills
- Confirm corrective action before accepting the next QC run
- Decide which patient results need review after rejected QC
- Set a new control lot's mean from its crossover runs and its SD from the established CV
At the bench: Review QC before releasing results
Try first
Get the idea
After a rejected QC run, patient results for that analyte are not reported until the problem is corrected. Three steps follow:
- Correct the fault.
- Confirm the correction with QC.
- Review the results already reported.
Correct the cause and confirm with QC
The investigation identifies a cause, such as:
- a control vial reconstituted incorrectly
- a reagent pack past its onboard stability
- a drifted calibration
- a clogged probe
The laboratory professional corrects that cause and then runs QC again. The test returns to service only after acceptable QC following a documented correction.
Rerunning the same controls until they pass skips the correction. The rerun may fall within the limits by chance with the fault still present, and the next run fails the same way.
CLIA requires the laboratory to document every corrective action it takes when control or calibration results fail its acceptance criteria, so the record shows what went wrong and what was done.1
Patient result lookback
- Patient results obtained in the unacceptable run, and all results reported since the last acceptable run, are evaluated to see whether they were affected.1
- The review goes back that far because a shift or trend can start before the run that violates a rule.
- Following the laboratory's procedure, specimens from the window are retested after the correction.
- When a reported result proves to be in error, the laboratory notifies the ordering provider and issues a corrected report promptly.3
A new control lot
Correction sometimes means opening a new control lot, and every new lot needs its own limits.
Establish the target mean from the new lot:
- The new lot has its own concentration, so its target mean comes from its own crossover runs, tested in parallel with the current lot. About 10 measurements on separate days give a starting target.2
- The old lot's mean would make every new-lot result look shifted.
- The insert mean is assigned across many laboratories and can differ from this analyzer's own mean.
Set the SD:
- A few days of crossover runs show only short-term scatter and give limits that are too tight.
- When the CV is steady across concentrations, the established long-term CV is applied to the new lot:2
new SD = new mean × established CV ÷ 100
The mean and SD are then updated as stable data accumulate.
Follow the bench demonstration
Sources for this step
References
- Electronic Code of Federal Regulations. 42 CFR §493.1282: Standard: Corrective actions. Accessed September 23, 2026.
- Clinical and Laboratory Standards Institute. Statistical Quality Control for Quantitative Measurement Procedures: Principles and Definitions. 4th ed. CLSI guideline C24Ed4E. Clinical and Laboratory Standards Institute; 2016. Accessed September 23, 2026.
- Electronic Code of Federal Regulations. 42 CFR §493.1291: Standard: Test report. Accessed September 23, 2026.
Watch one
At 07:00 both alanine aminotransferase (ALT) control levels read below −2 SD, violating the 2-2s rule.
- The last acceptable QC was at 15:00 yesterday.
- The reagent pack on board was opened 31 days ago, and its onboard stability is 28 days.
What has to happen before ALT results are reported again?
- Stop reporting ALT and keep the run's patient results pending release.
Results from a rejected run cannot be reported while their accuracy is in question.
- Take the expired reagent pack off and load a fresh pack.
Both levels moving the same way suggests a systematic cause, and a pack past its onboard stability is one.
- Run both control levels on the fresh pack and confirm they are inside their limits.
Only acceptable control results show that the new pack corrected the problem.
- Document the rejection, the expired pack, the replacement and the acceptable QC.
The record names the cause and the correction so it can be reviewed later.
- Evaluate the patient ALT results from the 07:00 run and every ALT reported since 15:00 yesterday.
A drift in reagent can start before the run that finally fails.
Your turn
Use it
Optional for this step.
Review missed questions
For reference
Sources checked
The rest of this step has a short briefing, a demonstration at the bench, 3 practice problems and a case in the lab.