Chromatography, electrophoresis, mass spectrometry and osmometry
15 min
Skills you'll build4 skills
- Predict elution order from how compounds interact with the stationary phase
- Predict protein migration from buffer pH relative to the isoelectric point
- Check the internal-standard response before releasing an LC-MS result
- Choose an osmometry principle that detects volatile solutes
At the bench: Review a separation pattern and its method controls · Review electrolyte results and related calculations for consistency
Try first
Get the idea
Chromatography
Each compound partitions between a moving mobile phase and a fixed stationary phase. The more a compound interacts with the stationary phase, the later it elutes.1,2
- Reversed phase. A nonpolar bonded phase, commonly C18, retains nonpolar compounds longest. Polar compounds elute first.
- Size exclusion. Large molecules cannot enter the pores of the packing, so they elute first.
A retention time or a thin-layer Rf supports identity only within one defined system and against standards run under the same conditions. Two compounds can coelute. A second solvent system, a color reaction or mass spectrometry supplies the confirming evidence.1,2
Electrophoresis and net charge
Buffer pH determines the sign of a protein's charge relative to its isoelectric point (pI).1,2
- Above its pI, a protein carries a net negative charge and moves toward the anode.
- Below its pI, it carries a net positive charge and moves toward the cathode.
- At its pI, net movement approaches zero.
Serum protein electrophoresis commonly runs at pH 8.6, above the pI of most serum proteins, so they move toward the anode.1 Electroendosmosis moves buffer toward the cathode and can carry weakly charged proteins with it.2
The internal standard in LC-MS
Quantitative liquid chromatography-mass spectrometry (LC-MS) adds a stable-isotope-labeled internal standard to each specimen early in preparation. The result comes from the ratio of analyte response to internal-standard response. That ratio corrects losses in extraction, injection and ionization to the extent the two compounds behave alike.3
The internal-standard response has its own acceptance limits. A sharp drop in one specimen suggests ion suppression or a preparation failure that the ratio may not fully correct. Re-extract or dilute that specimen and run it again before any result is released.2,3
Osmometry
Freezing-point depression responds to every dissolved particle. Each mOsm/kg lowers the freezing point by 0.00186 °C.1 Ethanol is volatile and enters the vapor phase, so it adds little to a vapor-pressure (dew-point) result. When an osmolal gap is calculated to look for a volatile solute such as ethanol, measure osmolality by freezing point, and keep the specimen capped so the ethanol does not evaporate.1,2
Follow the bench demonstration
Sources for this step
References
- Bishop ML, Fody EP, Van Siclen C, Mistler JM, Moy M. Clinical Chemistry: Principles, Techniques, and Correlations. 9th ed. Jones & Bartlett Learning; 2023.
- Rifai N, Chiu RWK, Young I, Burnham CAD, Wittwer CT, eds. Tietz Textbook of Laboratory Medicine. 7th ed. Elsevier; 2023.
- Clinical and Laboratory Standards Institute. Liquid Chromatography-Mass Spectrometry Methods. 3rd ed. CLSI guideline C62. Clinical and Laboratory Standards Institute; 2026. Accessed September 26, 2026. https://clsi.org/shop/standards/c62/
Watch one
The emergency department sends serum for measured osmolality so that an osmolal gap can be calculated. The patient is suspected of drinking a large amount of alcohol. The tube arrives capped and full. Your laboratory has a vapor-pressure osmometer on the main line and a freezing-point osmometer in the stat area. Which instrument do you use, and how do you handle the tube?
- Identify the purpose: the gap is calculated to find unmeasured solutes, and ethanol is one of them.
The purpose of the test determines the measurement principle.
- Rule out the vapor-pressure osmometer: it would miss most of ethanol's contribution to the measured osmolality and underestimate the gap.
Ethanol enters the vapor phase and adds little to a dew-point reading.
- Choose the freezing-point osmometer: it measures ethanol along with every other dissolved particle.
Freezing-point depression responds to particle number, regardless of the particle.
- Keep the tube capped until the sample is taken, and analyze it promptly.
Ethanol that evaporates from an open tube is missing from the result.
Your turn
Use it
Optional for this step.
- A night batch of 40 whole-blood tacrolimus specimens runs by LC-MS with tandem detection.
- Calibrators and both control levels are within limits. Every other specimen's internal standard is within limits.
- The procedure accepts an internal-standard peak area from 50% to 150% of the mean area of the run's calibrators.
- One transplant patient's specimen (MRN 0072915) gives the results below.
| Test | Result | Previous | Reference interval | Flag |
|---|---|---|---|---|
| Tacrolimus | 12.9 ng/mLSep 24, 06:05: 7.8 ng/mL | 7.8 ng/mLSep 24, 06:05 | ||
| Internal-standard area | 18 % LowSep 24, 06:05: 97 % | 97 %Sep 24, 06:05 | 50–150 %50–150 | Low |
Specimen: Not measured on whole blood. Whole blood, EDTA. Collected 05:50 before the morning dose.
The internal-standard area of 18% stands out in a run where everything else passed. The problem is limited to this specimen. A fresh extraction with an acceptable internal standard gives the reportable result.
Review missed questions
For reference
Sources checked
The rest of this step has a short briefing, a demonstration at the bench, 3 practice problems and a short case.