Extraction, purity, PCR formats and controls
15 min
Skills you'll build4 skills
- Choose an extraction method for a DNA, RNA, or cell-free nucleic acid target
- Accept or reject a nucleic-acid extract using A260/A280, integrity, and amplifiability
- Choose endpoint, real-time, digital, or RT-PCR for the target and for quantitation
- Locate the failed step of a molecular test from where each control is added
At the bench: Prepare a molecular run with the required controls
Try first
Get the idea
Choosing an extraction method
Every extraction lyses the specimen and removes substances that inhibit or interfere with the next step. How much target it recovers depends on the matrix, where the target is located, its length, the input and the elution volume.1 A method that recovers plenty of genomic DNA from white cells can lose short cell-free DNA fragments or RNA. RNA testing also requires control of ribonuclease exposure and adds a reverse-transcription step before amplification.1,2
Purity, integrity and amplifiability
Absorbance at 260 nm estimates total nucleic acid. The A260/A280 ratio of pure DNA is near 1.8, and pure RNA is near 2.0.3 A low ratio suggests gross protein or solvent contamination. An acceptable ratio gives no information about fragment size, degradation or inhibitors. Electrophoresis shows size and degradation, and an amplification control shows whether the extract amplifies. The assay's own acceptance limits determine whether an extract can be used.1
PCR formats
| Format | What it measures well |
|---|---|
| End-point PCR | Presence of product after cycling |
| Reverse-transcription PCR | An RNA target, copied to DNA first |
| Real-time PCR | Starting quantity, read from the quantification cycle (Cq) against a standard curve |
| Digital PCR | Starting quantity, counted as positive partitions |
Reactions that start with very different amounts of target reach similar signals at the plateau. End-point signal therefore cannot rank starting quantity.2 A lower Cq means more amplifiable target, and Cq values can be compared only across runs with compatible efficiency, threshold and calibration.4
PCR controls
- Negative extraction control: enters before extraction and detects contamination from extraction onward.
- No-template control: enters at reaction setup and detects contaminated reagents or setup.
- Positive control: detects failure to find a known target, over the steps after the point where it is added.
- Internal amplification control: travels with each specimen and detects inhibition or reaction failure.2,5
A positive control added after extraction cannot show that extraction worked. When a control fails, the failed step lies between the point where that control entered and the end of the run.5
Follow the bench demonstration
Sources for this step
References
- Clinical and Laboratory Standards Institute. Collection, Transport, Preparation, and Storage of Specimens for Molecular Methods. 2nd ed. CLSI guideline MM13. Clinical and Laboratory Standards Institute; 2020. Accessed September 26, 2026.
- Kralik P, Ricchi M. A basic guide to real time PCR in microbial diagnostics: definitions, parameters, and everything. Front Microbiol. 2017;8:108. doi:10.3389/fmicb.2017.00108
- Powell EA, Mortensen JE. Extraction of total nucleic acids from bacterial isolates using the bioMérieux NucliSENS easyMAG total nucleic acid extractor. Ann Clin Microbiol Antimicrob. 2016;15(1):54. doi:10.1186/s12941-016-0168-7
- Bustin SA, Ruijter JM, van den Hoff MJB, et al. MIQE 2.0: revision of the Minimum Information for Publication of Quantitative Real-Time PCR Experiments guidelines. Clin Chem. 2025;71(6):634-651. doi:10.1093/clinchem/hvaf043
- Clinical and Laboratory Standards Institute. Molecular Diagnostic Methods for Infectious Diseases. 3rd ed. CLSI report MM03. Clinical and Laboratory Standards Institute; 2015. Accessed September 26, 2026.
Watch one
A DNA extract from a blood specimen is ready for an assay whose target is 450 base pairs long.
- Concentration is 42 ng/µL, above the assay's minimum input.
- A260/A280 is 1.84.
- Capillary electrophoresis shows most fragments below 300 base pairs, with little intact high-molecular-weight DNA.
Do you accept the extract for the assay?
- Read the ratio: 1.84 is near the 1.8 expected for DNA, so gross contamination is unlikely.
The ratio is the quickest screen for gross protein or solvent contamination.
- Look at integrity next: most fragments are shorter than 300 base pairs.
An acceptable ratio gives no information about fragment size or degradation.
- Compare with the target: a 450-base-pair target needs fragments at least that long, and few are present.
The assay can amplify only a template that spans its whole target.
- Compare with the assay's acceptance limits: this extract fails the integrity requirement.
Degraded input amplifies poorly and can produce a false negative.
- Re-extract from the specimen under the procedure, or request a new specimen if the specimen itself is degraded.
A new extraction shows whether the degradation came from the extract or the specimen.
Your turn
Use it
Optional for this step.
- A reverse-transcription real-time PCR run for a respiratory RNA virus includes 22 patient swabs.
- The negative extraction control went through extraction with the patients and shows the viral target at Cq 35.1.
- The no-template control, added at reaction setup, shows no amplification.
- The positive control is detected at its expected Cq.
- Every internal control is detected in range.
- Three patient swabs show the viral target at Cq 34.6 to 36.8.
- The run's patient results are pending release.
The run is judged by which negative control detected the target. The negative extraction control went through extraction and was positive. The no-template control was added later and was negative. The contamination therefore occurred during extraction, so the whole run is repeated from extraction with the method the assay was validated with.
Review missed questions
For reference
Sources checked
The rest of this step has a short briefing, a demonstration at the bench, 3 practice problems and a short case.
Free account: keep your progress and get review reminders. Have one? Sign in