Pipettes, glassware and temperature checks
15 min
Skills you'll build4 skills
- Verify temperature at the working sample position
- Follow the to-contain or to-deliver marking on volumetric equipment
- Choose a pipette type for the volume and for viscous, volatile, or foaming liquids
- Calculate delivered volume from water mass and the stated conversion factor
At the bench: Perform and document required equipment checks
Try first
Get the idea
Checking temperature at the sample position
A water bath or heating block display reports the controller's own sensor. The temperature at the specimen can differ with position, gradients and drift, so the check is made at the sample position.1
- Let the bath and a calibrated reference probe equilibrate.
- Read the probe at the working setpoint in representative rack positions.
- Apply the probe's calibration correction.
- Compare each corrected reading with the procedure's acceptance interval.1
The probe's calibration record names the reference device, the correction factors and their uncertainty. A "NIST traceable" label does not document traceability without that record.2
Pipette drainage markings
To-contain (TC) ware holds its stated volume. To-deliver (TD) ware releases its stated volume when its drainage procedure is followed.3
- A continuous etched band or double ring near the top marks a blowout pipette. Its last drop is expelled.
- A pipette without that mark is self-draining. The drop left in its tip is part of its calibration.1
Using the wrong drainage rule adds or leaves behind the same volume on every transfer.
Choosing a pipette for the liquid
An air-displacement pipette moves liquid through an air cushion. Viscosity, volatility, foaming and temperature differences change the volume delivered. A positive-displacement pipette moves its piston in contact with the liquid and handles these liquids better. Reverse pipetting is another option. Either technique is verified locally. The pipette's operating range must cover the volume.1
Gravimetric calibration
Gravimetric calibration weighs replicate deliveries of purified water. The factor Z converts mass to volume and corrects for water density and air buoyancy.4
mean delivered volume = mean mass × Z
systematic error = mean delivered volume − selected volume
At 22.0 °C and 101.3 kPa, Z is 1.0033 µL/mg.4 A 100 µL pipette with a mean mass of 99.29 mg delivers 99.29 mg × 1.0033 µL/mg = 99.62 µL, a systematic error of −0.38 µL. Reading the milligrams as microliters overstates the error as −0.71 µL. The laboratory compares the systematic and random errors with its acceptance limits.5
Follow the bench demonstration
Sources for this step
References
- Bishop ML, Fody EP, Van Siclen C, Mistler JM, Moy M. Clinical Chemistry: Principles, Techniques, and Correlations. 9th ed. Jones & Bartlett Learning; 2023.
- Bruce SS, Possolo A, Watters RL Jr. Metrological Traceability: Frequently Asked Questions and NIST Policy. NIST Technical Note 2156. National Institute of Standards and Technology; 2021. Accessed September 26, 2026.
- ASTM International. Standard Specification for Laboratory Glass Volumetric Apparatus. ASTM E694-18(2024). ASTM International; 2024.
- International Organization for Standardization. Piston-operated volumetric apparatus — Part 6: Gravimetric reference measurement procedure for the determination of volume. ISO 8655-6:2022. International Organization for Standardization; 2022. Accessed September 26, 2026.
- International Organization for Standardization. Piston-operated volumetric apparatus — Part 2: Pipettes. ISO 8655-2:2022. International Organization for Standardization; 2022. Accessed September 26, 2026.
Watch one
You check a 500 µL air-displacement pipette at its nominal setting.
- Ten deliveries of purified water weigh 4,974.0 mg in total.
- The water and air are at 22.0 °C and the pressure is 101.3 kPa, so Z is 1.0033 µL/mg.
- The laboratory's procedure accepts a systematic error within ±4.0 µL at this setting.
Does the pipette pass for systematic error?
- Find the mean mass: 4,974.0 mg ÷ 10 = 497.40 mg.
The check uses the mean of the replicate deliveries.
- Convert to volume: 497.40 mg × 1.0033 µL/mg = 499.04 µL.
Z corrects for water density and air buoyancy, so each milligram on the balance corresponds to slightly more than 1 µL.
- Subtract the selected volume: 499.04 µL − 500 µL = −0.96 µL, which is −0.19% of the setting.
The sign shows the direction, and a negative error means the pipette delivers too little.
- Compare with the limit: −0.96 µL lies within ±4.0 µL.
The procedure's acceptance limit determines pass or fail for this setting.
Your turn
Use it
Optional for this step.
- You set up a manual enzyme immunoassay on the special chemistry bench.
- Step 1 adds 20 µL of a viscous, glycerol-based conjugate to each tube.
- Step 2 incubates the tubes for 30 min in a heating block set to 37 °C. The procedure accepts 36.0 to 38.0 °C in every well used.
- Step 3 adds 1.0 mL of stop solution with a 1 mL serologic pipette marked TD, with a double ring near the mouthpiece.
- This morning's check with a calibrated reference probe, correction applied, read 37.1 °C in a center well and 35.6 °C in a corner well. The display reads 37.0 °C.
Each decision in this setup came from something you could check:
- The conjugate is viscous, so it needs a positive-displacement pipette.
- The corner well read 35.6 °C after correction, below the 36.0 °C limit, so the heating block failed its check. The display and the center well did not show the problem.
- The double ring near the mouthpiece marks the serologic pipette as a blowout pipette.