Skip to content
LatticeMLS, Home

Add your earlier progress to your account?

From before you signed in, this browser has:

Which plan do you want to keep?

You changed your plan before signing in, and those changes don't match the plan saved on your account.

SettingSaved planPlan changes

Study progress is waiting to be saved

Water, vessels, balances and centrifuges

15 min

Skills you'll build4 skills
  • Choose a vessel and cleaning process that will not alter the analyte
  • Tell balance readability from repeatability measured with mass standards
  • Calculate relative centrifugal force from radius and rotational speed
  • Choose resistivity, organic-carbon, or plate-count checks for the water an assay needs

At the bench: Perform and document required equipment checks

Try first

Try first

A serum specimen for a zinc level needs an aliquot sent to a reference laboratory for trace-metal testing. Which container do you use?

Keep going to see the reasoning.

Keep going to see the reasoning.

Right. Keep going to follow the reasoning.

Keep going to see the reasoning.

Get the idea

Container compatibility

A container can adsorb an analyte, release additives, or let gas and vapor through. The effect depends on the material, the analyte, the contact time and the temperature.1 Glass adsorbs metal ions, so ordinary glass is a poor container for trace-metal work. Labels such as certified-clean or low-binding are separate product properties, each checked against its own specification.

Rinse washed ware repeatedly with water of the required grade. Compare the pH of the final rinse with the pH of the incoming rinse water. An alkaline drift means detergent is left behind. Residue can shift pH, suppress enzymes and add optical background.1

Readability and repeatability

Readability is the smallest step the balance displays. Repeatability is how closely repeated weighings of one load agree. Linearity is the response across the range. A balance that reads to 0.1 mg can scatter far more than that when drafts, vibration, static or poor leveling disturb it.1 Checks with traceable mass standards at masses across the working range show how well it performs.2,3

Relative centrifugal force

RCF = 1.118 × 10−5 × r × rpm2

Here r is the radius in centimeters from the axis of rotation to the point in the tube the procedure names. The rotor manual lists it for each tube position. Radius enters the formula once, and speed enters as its square. The same rpm in a rotor with a different radius gives a different force, so program the RCF, or convert it to rpm with each rotor's own radius.1

Laboratory reagent water

CLSI GP40 classifies laboratory water by intended use, and the laboratory validates that the water it selects is suitable for each purpose. Each monitored parameter detects one kind of contamination.4

CheckDetects
ResistivityIonic contamination
Total organic carbonOrganic contamination
Viable plate countMicrobial burden

Storage and distribution can recontaminate purified water, so monitoring samples are taken at the point of use.4

Follow the bench demonstration

Sources for this step
References
  1. Bishop ML, Fody EP, Van Siclen C, Mistler JM, Moy M. Clinical Chemistry: Principles, Techniques, and Correlations. 9th ed. Jones & Bartlett Learning; 2023.
  2. ASTM International. Standard Specification for Laboratory Weights and Precision Mass Standards. ASTM E617-23. ASTM International; 2023.
  3. Bruce SS, Possolo A, Watters RL Jr. Metrological Traceability: Frequently Asked Questions and NIST Policy. NIST Technical Note 2156. National Institute of Standards and Technology; 2021. Accessed September 26, 2026.
  4. Clinical and Laboratory Standards Institute. Preparation and Testing of Reagent Water in the Medical Laboratory. 5th ed. CLSI guideline GP40. Clinical and Laboratory Standards Institute; 2024. Accessed September 26, 2026.

Watch one

The serum procedure specifies 1,400 × g. The swinging-bucket rotor's manual lists a maximum radius of 16.0 cm, and the procedure names the maximum radius. The centrifuge's run card says 2,800 rpm. Does 2,800 rpm give the required force in this rotor?

  1. Take the radius: r = 16.0 cm.

    The procedure says which radius applies, and the manual gives its value for this rotor.

  2. Square the speed: 2,8002 = 7,840,000.

    Speed enters the formula as its square, so it is calculated first.

  3. Calculate: RCF = 1.118 × 10−5 × 16.0 × 7,840,000 = 1,402 × g (rounded).

    The constant and the radius multiply the squared speed.

  4. Compare: 1,402 × g meets the specified 1,400 × g.

    The procedure names a force, so the calculated force is compared with it.

Yes. In this rotor 2,800 rpm gives about 1,402 × g, which meets the procedure's 1,400 × g.

Practice this skill

Your turn

Problem 1 of 3 · Hints available

Two rotors spin at the same rpm. One has an effective radius of 5 cm and the other 10 cm. How does the RCF at 10 cm compare with the RCF at 5 cm?

Incorrect. RCF depends on radius as well as speed, so the same rpm gives a different force in a rotor of a different radius.

Correct. In RCF = 1.118 × 10⁻⁵ × r × rpm², radius is a direct multiplier, so going from 5 cm to 10 cm at the same rpm doubles RCF.

Incorrect. Only the speed term is squared. Radius enters the formula once, so doubling it doubles RCF, and doubling rpm would quadruple it.

Hint
  1. Look at which term in the formula is squared.
  2. Hold rpm fixed and see what happens to the product when only r changes.

Review Centrifugation

Problem 2 of 3 · Hints available

An analytical balance reads to 0.1 mg. The procedure requires a repeatability standard deviation of 0.2 mg or less. Ten weighings of a 20 g mass standard give a standard deviation of 0.6 mg. The balance stands beside an open door to the corridor. What do you conclude?

Readability is only the size of the last displayed step. The repeated weighings scatter with a standard deviation of 0.6 mg, three times the procedure's limit.

Took balance readability as its weighing accuracy

Readability is the smallest increment the balance displays. Drafts, vibration, static, poor leveling, or nonlinearity can cause bias or imprecision far larger than that increment. Repeatability and working-range checks with traceable mass standards show the balance's actual weighing performance.

Checks use standards at masses across the working range, and 20 g lies well within an analytical balance's capacity. A lighter standard would not remove the scatter.

A standard deviation of 0.6 mg exceeds the 0.2 mg limit. Drafts from the open door are a likely cause, and a repeat check after the cause is corrected shows whether the balance is fit for use.

Hint
  1. Readability and repeatability measure different things.
  2. Compare the measured standard deviation with the procedure's limit.
  3. Think about what an open door does to the air around the pan.

Review Balances, desiccators, and mass standards

Problem 3 of 3 · On your own

A fixed-angle rotor has an effective radius of 15.0 cm at the point the procedure names. It spins at 2,500 rpm. What is the relative centrifugal force, to the nearest whole number?

Show the answer

1,048 × g

RCF = 1.118 × 10−5 × 15.0 × 2,5002 = 1.118 × 10−5 × 15.0 × 6,250,000 = 1,048 × g.

Review Centrifugation

Use it

Optional for this step.

  • The chemistry section is preparing reagents for a new, sensitive enzymatic assay. Its instructions require water low in organic carbon and microbes.
  • The water system's point-of-use resistivity monitor reads 18.0 MΩ·cm, within its limit.
  • The reagent glassware comes from the washer. The final rinse reads pH 8.4, and the incoming rinse water reads pH 6.8.
  • A new centrifuge arrives with a swinging-bucket rotor whose radius at the point the procedure names is 18.0 cm. The serum procedure specifies 1,300 × g for 10 min. The old rotor, 13.0 cm, ran it at 3,000 rpm.
Decision 1 of 3

Which evidence shows that the water is suitable for the new assay?

Resistivity measures ionic contamination only. Water with high resistivity can still contain organic compounds and microorganisms, which the assay's instructions limit.

Took high resistivity to mean all contaminants were absent

Resistivity measures ionic contamination only. Water with high resistivity can contain organic compounds or microorganisms, which total organic carbon testing and plate counts detect, and these can interfere with sensitive assays. Monitor at the point of use to check the water the assay receives.

Storage and distribution can recontaminate water after the purifier. A sample from the point of use shows the water the assay receives.

Total organic carbon detects organic contamination and plate counts detect microbes, the two things the instructions limit. Sampling at the point of use tests the water the assay receives.

Review Reagents and laboratory water

Decision 2 of 3

What do you do with the washed glassware?

A final rinse at pH 8.4 against incoming water at pH 6.8 is an alkaline drift, which indicates detergent residue on the glass. Residue can suppress the enzyme and add optical background, so the glass is rinsed until the drift is gone.

Detergent residue is invisible. The rinse pH shows it is there, and it can shift pH, suppress enzymes and add optical background in the assay.

Oven drying leaves the detergent on the glass. Rinsing with water of the required grade removes it.

Review Containers, cleaning, and compatibility

Decision 3 of 3

What do you program for the serum spin on the new centrifuge?

At 18.0 cm, 3,000 rpm gives 1.118 × 10−5 × 18.0 × 3,0002 = 1,811 × g, about 40% more force than the procedure specifies.

Copied rpm to a rotor with a different radius

RCF depends on radius as well as rpm, so the same rpm in a rotor of different radius gives a different force. Copying a speed between rotors applies too little or too much force and changes the separation. Convert the procedure's RCF to rpm with each rotor's own radius.

A larger radius already gives more force at the same speed, so the speed must be lower. Scaling it up gives a force far above 1,300 × g.

Programming the procedure's force gives the correct force in any rotor. Converted with this rotor's 18.0 cm radius, 1.118 × 10−5 × 18.0 × 2,5402 = 1,298 × g.

Review Centrifugation

Each decision used the measurement that fits the problem:

  • Point-of-use organic carbon and plate counts show whether water meets an assay's limits for organics and microbes. Resistivity measures only ionic contamination.
  • A final rinse more alkaline than the incoming water shows detergent left on the glass.
  • The new rotor's 18.0 cm radius determines the speed. The old rotor's 3,000 rpm gave 1,308 × g because its radius was 13.0 cm.

For reference

Sources checked